Model Number 0570-0404
Medical Device Problem Codes Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer declined the option to have their glidescope core 15-inch monitor returned to verathon for evaluation.Since the device was not returned to verathon for evaluation, the cause could not be determined.Upon verathon following up with the customer for further troubleshooting, the customer reported being unable to isolate the issue to a particular system component and declined the option to have their system returned to verathon for evaluation.No information was provided regarding the cable or video laryngoscope used with the monitor during the reported event.Review of complaint history for the reported monitor serial number (b)(6) did not identify any previous complaints reported to verathon.Trending analysis for glidescope core monitors does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event or Problem Description
A customer reported that during a patient procedure, using a glidescope core 15-inch monitor, the screen abruptly went black twice.It was reported that the image was restored after the cable was wiggled.Following the reported incident, the customer reported being unable to duplicate the problem or isolate the issue to a particular system component.No delay in the procedure, use of a backup device, or harm to the patient was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE CORE 15-INCH MONITOR
Common Device Name
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key18565010
Report Number9615393-2024-00010
Device Sequence Number12793111
Product Code CCW
UDI-Device Identifier00879123006639
UDI-Public010087912300663911200602
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0570-0404
Device Catalogue Number0270-1033
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/28/2023
Initial Report FDA Received Date01/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/18/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-