• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXIMA; TUBE, DRAINAGE, SUPRAPUBIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXIMA; TUBE, DRAINAGE, SUPRAPUBIC Back to Search Results
Model Number 32375
Medical Device Problem Codes Unsealed Device Packaging (1444); Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that device packaging seal was compromised.A 10.3 flexima drainage catheter was selected for use.However, it was noted that there was a hole in the packaging, and the seal was compromised.There was no patient or procedure involved.
 
Event or Problem Description
It was reported that device packaging seal was compromised.A 10.3 flexima drainage catheter was selected for use.However, it was noted that there was a hole in the packaging, and the seal was compromised.There was no patient or procedure involved.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the complaint device was received for product analysis.It is possible to observed that the package was ripped/torn, wrinkles were observed at the torn section and with waste/residues were inside the package at torn section.Also, the photo attached on the complaint shows a hole on the device pouch.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that device packaging seal was compromised.A 10.3 flexima drainage catheter was selected for use.However, it was noted that there was a hole in the packaging, and the seal was compromised.There was no patient or procedure involved.
 
Additional Manufacturer Narrative
Device eval by mfr: the device was not returned.However, a photo was provided from the healthcare facility.It shows a hole on the device pouch.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIMA
Common Device Name
TUBE, DRAINAGE, SUPRAPUBIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18565438
Report Number2124215-2024-01634
Device Sequence Number17845297
Product Code GBO
UDI-Device Identifier08714729323310
UDI-Public08714729323310
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/09/2026
Device Model Number32375
Device Catalogue Number32375
Device Lot Number0032191536
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/04/2024
Supplement Date Received by Manufacturer02/07/2024
03/06/2024
Initial Report FDA Received Date01/23/2024
Supplement Report FDA Received Date02/22/2024
03/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-