• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL BIRD¿ LOW FLOW AIR/O2BLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL BIRD¿ LOW FLOW AIR/O2BLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION Back to Search Results
Model Number BIRD¿ LOW FLOW AIR/O2BLENDER
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/02/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Vyaire medical file identification: (b)(4).H3: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event or Problem Description
It was reported to vyaire medical that the bird¿ low flow air/o2blender has a problem with the knob (has not limit) and needs overhaul.As of this time, there is no information about patient involvement associated with this reported event.
 
Additional Manufacturer Narrative
Result of investigation: the reported failure is not verified.Found bent stem valve.Overhaul and calibration performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIRD¿ LOW FLOW AIR/O2BLENDER
Common Device Name
MIXER, BREATHING GASES, ANESTHESIA INHALATION
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
sandra valencia
510 technology drive
irvine, CA 92618
8477865998
MDR Report Key18567719
Report Number2021710-2024-18687
Device Sequence Number8518005
Product Code BZR
UDI-Device Identifier10846446002355
UDI-Public(01)10846446002355
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K883038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBIRD¿ LOW FLOW AIR/O2BLENDER
Device Catalogue Number03920A
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/02/2024
Supplement Date Received by Manufacturer05/17/2024
Initial Report FDA Received Date01/23/2024
Supplement Report FDA Received Date06/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-