• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN MULTIHOLE W/GRIPTION 54MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN MULTIHOLE W/GRIPTION 54MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 121730054
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 01/15/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Patient had experienced multiple dislocations post tha.Surgeon attempted to convert to dual mobility but hip was still unstable.Cup was removed and a new cup was placed with appropriate anteversion and a dual mobility articulation without patient harm or delay.Doi: (b)(6) 2021.Dor: (b)(6) 2024.Affected side: left hip.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot the device lot number is unknown, therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.H6: component code: appropriate term/code not available (g07002) used to capture no findings available.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MULTIHOLE W/GRIPTION 54MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18569919
Report Number1818910-2024-01606
Device Sequence Number8491851
Product Code KWA
UDI-Device Identifier10603295009993
UDI-Public10603295009993
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K093646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number121730054
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/16/2024
Supplement Date Received by Manufacturer01/24/2024
Initial Report FDA Received Date01/24/2024
Supplement Report FDA Received Date01/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX +4 10D 36IDX54OD.; DELTA CER HEAD 12/14 36MM +8.5.; PINN CAN BONE SCREW 6.5MMX25MM.; PINN CAN BONE SCREW 6.5MMX25MM.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexMale
Patient Weight94 KG
-
-