| Lot Number |
0032263714 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The target lesion was located in a mildly tortuous and moderately calcified coronary artery.A 4.00 x 28mm synergy xd drug eluting stent was advanced for treatment.However, it was noted that the shaft broke during insertion.The device was removed and completed the procedure with another synergy stent.There were no patient complications reported.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The target lesion was located in a mildly tortuous and moderately calcified coronary artery.A 4.00 x 28mm synergy xd drug eluting stent was advanced for treatment.However, it was noted that the shaft broke during insertion.The device was removed and completed the procedure with another synergy stent.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: synergy xd mr us 4.00 x 28mm stent delivery system was returned for analysis.Visual, tactile and microscopic analysis were performed on the device.A visual and tactile examination of the hypotube found multiple kinks along the shaft and also a shaft break at 26cm distal to the distal end of the strain relief.No issues identified with the outer / mid-shaft sections or the inner lumen of the device.Examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.Balloon cones were reviewed, and no issues were noted.Balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.No signs of movement, stent was set between the proximal and distal markerbands.Bumper tip showed no signs of distal tip damage.No other device issues were identified during returned product analysis.
|
| |
|
Search Alerts/Recalls
|