|
AOMORI OLYMPUS CO., LTD. SINGLE USE 3-LUMEN SPHINCTEROTOME V; SINGLE USE 2-LUMEN SPHINCTEROTOME
|
Back to Search Results |
|
| Model Number |
KD-V411M-0720 |
| Medical Device Problem Codes |
Break (1069); Melted (1385); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus and evaluated.The customer¿s allegation was confirmed and additionally the knife wire coating was found to be torn and the broken portion was burnt and melted.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus that during a therapeutic endoscopic sphincterotomy the knife wire of the single use 3-lumen sphincterotome v broke when it was energized.The customer reported there was no shedding inside the patient¿s body.The high frequency equipment and settings used were not specified.The procedure was completed using a similar device.There was no patient injury or medical intervention associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation a definitive cause for the broken knife wire was not established, however the following factors may have contributed to the reported event including: 1.The device was not protruded enough from the endoscope until the rear end of the cutting wire was in the field of view.2.Due to the situation of ¿1¿ description, the cutting wire and the endoscope were being close to each other.3.The output was activated in state of ¿2¿ description.This might have led to an electrical discharge between the cutting wire and the distal end of the endoscope.4.An electrical discharge possibly occurred, and the cutting wire became hot instantly.That might have caused the cutting wire to break.5.Since the cutting wire was broken, it was felt that the cutting wire was no longer energized.The occurrence of the reported problem can be prevented by adhering to the instructions for use (ifu) which states the following: ¿¿since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.¿ when inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.¿do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.¿if you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion.¿be sure that the rear end of the cutting wire is extended from the distal end of the endoscope.Insufficient output or unintended tissue injury may occur in case the cutting wire contacts the forceps elevator.¿ olympus will continue to monitor the field performance of this device.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation it is possible that the slider was slightly pushed causing the cutting wire to deflect.The coated portion of the cutting wire may have come into contact with metal portion of device causing peeling.Additionally, physical stress may have applied to the coated portion of the cutting wire when the device was removed from endoscope after the coated portion of the cutting wire has torn.This likely caused the coated portion of the cutting wire to detach from the cutting wire.As a result, the coated portion was possibly partially missing.The occurrence of the reported problem can be prevented by adhering to the instructions for use (ifu) which states the following: ¿since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.When inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.Do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.If you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion.Be sure that the rear end of the cutting wire is extended from the distal end of the endoscope.Insufficient output or unintended tissue injury may occur in case the cutting wire contacts the forceps elevator.¿.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|
|
|