| Model Number |
PUNCH, WIDEBITER, 3.4 MM STRAIGHT |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 28-dec-2023, it was reported by a sales representative via (b)(4) that a widebiter punch, ar-12040, split into two pieces.Occurred during a case, no adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.One unpacked ar-12040 serial/batch number (b)(6) was received for evaluation.Upon visual inspection, it was determined that the finger had become detached from the instrument, likely due to a missing pin.The most probable cause of the reported failure is misuse by the end user.
|
| |
|