| Catalog Number |
121730052 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/11/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Liner breakage intra-op & shell breakage intra-op.Patient underwent total hip arthroplasty.The ceramic liner was found to be fractured.At the same time, the shell was found to be damaged.Changed another shell and liner to complete the surgery.There was a 1 hour delay in the procedure.The patient was in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received: a.Can you please provide the original implantation date? (b)(6) 2024.B.Was there any patient harm related to the event? there was a 1 hour delay in the procedure.C.Kindly confirm the alert date of the complaint? when were you made aware of the event? (b)(6) 2024.D.Kindly verify the event date? 11th jan 2024.E.Did it break into two or more pieces? if yes, were all the fragments retrieved from the patient? yes, it break into more than two pieces.And the fragments retrieved from the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : according to the information received, ¿liner breakage intra-op & shell breakage intra-op.Patient underwent total hip arthroplasty.The ceramic liner was found to be fractured.At the same time, the shell was found to be damaged.Changed another shell and liner to complete the surgery.There was a 1 hour delay in the procedure.The patient was in stable condition now¿.Visual analysis of the returned sample revealed that the rim and the inner taper portion of the pinn multihole w/gription 52mm exhibits signs of damage and scratches.However, it cannot be determined whether this scratches occurred prior to the insertion of the ceramic liner or after the fracture event.No signs of fracture were observed on the acetabular cup.With the information provided is not possible to determine a potential cause at this moment.The mode of failure of the device is multi-factorial and consideration must be given to all other potential influences during the surgical process.Any conclusions from the investigational input provided have to be placed into context with all other relevant factors.The overall complaint was not confirmed as the observed condition of the pinn multihole w/gription 52mm would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a manufacturing record evaluation was performed for the finished device [121730052 / m30f16] number, and no non-conformances / manufacturing irregularities were identified during manufacturing.Device history review : a manufacturing record evaluation was performed for the finished device [121730052 / m30f16] number, and no non-conformances / manufacturing irregularities were identified during manufacturing.
|
| |
|