| Model Number |
BF-1T180 |
| Medical Device Problem Code |
Device Reprocessing Problem (1091)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not yet been returned to olympus for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus that the evis exera ii bronchovideoscope had leakage inside the instrument channel and the plastic distal end cover burnt or melted around the instrument channel at the distal end.The issue was found during a therapeutic biopsy procedure.There were no reports of patient harm.
|
| |
|
Event or Problem Description
|
|
The reported issue was found after the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the device evaluation and the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over nine (9) years since the subject device was manufactured.The device was returned to olympus for inspection, and the customer's reportable malfunction was confirmed.Based on the investigation results, it was found that during the use of an endo therapy accessory for high-frequency cauterization, leaked current flowed through the patient via the light guide lens end of the device, causing the flow of high-frequency current in the patient.From the investigation it is presume that the deterioration of the glue filled between the distal end and the plastic distal end cover/probe cover led to poor insulation performance.The event can be prevented by following the instructions for use which state: "-operation manual 3.2 inspection of the endoscope 9.Inspect the objective lens and light guide lens at the distal end of the endoscope¿s insertion section for scratches, cracks, stains, gaps around the lens, or other irregularities." olympus will continue to monitor field performance for this device.
|
| |
|