• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL COSTA RICA LTD. EXT SET 7IN CLV SL; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL COSTA RICA LTD. EXT SET 7IN CLV SL; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 126600490
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2023
Type of Reportable Event Malfunction
Event or Problem Description
The event involved an ext set 7in clv sl where it was reported that the tubing burst during contrast injection in ct scan.These loops were used for high pressure injection.The product was connected to patient and ct injector was connected to luer lock port.Due to the event the scans had poor images due to insufficient contrast.There was patient involvement and no patient harm.
 
Additional Manufacturer Narrative
The device was returned for evaluation; however, testing has not yet been completed.
 
Additional Manufacturer Narrative
Received one used 146870489 primary plum set for inspection.As received, the tubing on set was ruptured and flared.This type of tubing damage is consistent with a high pressure injection during use.The dfu states: caution: not for high pressure infusion.The reported complaint can be confirmed.The probable cause is due to a high pressure injection on a device that is not rated for high pressure during use.The lot history was reviewed, no nonconformities were identified that may have contributed to the reported complaint.
 
Event or Problem Description
Two devices were returned for evaluation.This captures the first of two devices returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXT SET 7IN CLV SL
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18572314
Report Number9615050-2024-00068
Device Sequence Number7140396
Product Code FPA
UDI-Device Identifier10887787005612
UDI-Public(01)10887787005612(17)260701(10)13639925
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K151969
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2026
Device Catalogue Number126600490
Device Lot Number13639925
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/26/2023
Supplement Date Received by Manufacturer02/21/2024
Initial Report FDA Received Date01/24/2024
Supplement Report FDA Received Date03/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CONTRAST INJECTOR, UNK MFR; CONTRAST, UNK MFR
Patient SexUnknown
-
-