• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTERVASCULAR SAS CARDIOROOT; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTERVASCULAR SAS CARDIOROOT; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER Back to Search Results
Model Number HEWROOT0030
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(10/3233) a fragment of the device was available and returned to intervascular.The examination of the fragment will be performed.(4109/213) the review of historical data indicated that no other similar complaint was reported for the same sterilization lot number 23c30.(3331/213) the device history records review concluded that no deviation was identified in relation with the reported event.(11) the investigation is still ongoing.A follow-up report will be sent upon completion of the investigation.
 
Event or Problem Description
It was reported to intervascular that the cardioroot was cut in two parts and when using the distal part of the prosthesis for wrapping, the surgeon noticed the presence of debris and intra-prosthetic fragments resembling collagen with embolism risk.The wrapping was done nevertheless.Additional information received from the initial reporter indicates that the issue was noticed during a tirone david procedure when the involved fragment was opened to wrap the distal (extra-vascular) anastomosis.No defects were observed on the implanted fragment.The surgery was not delayed and no consequences were reported on the patient.Moreover, the surgeon is not a new user of cardioroot.
 
Event or Problem Description
Complaint # (b)(4).
 
Additional Manufacturer Narrative
Corrected data : on block h1, type of reportable event was corrected from serious injury to malfunction.Indeed, according to the information received from the initial reporter the reported event does not meet the definition of a ¿serious injury¿ as described in the 21 cfr part 803.3.Actually, it did not result in life-threatening illness or injury or in permanent impairment of a body function or permanent damage to a body structure, or necessitated medical or surgical intervention to preclude permanent impairment of a body function or permanent damage to a body structure.(10/4248) a fragment of the involved device was returned to intervascular for examination.A visual inspection was performed by the quality assurance (qa) supervisor, the observation results are as follows: ¿collagen fragments were noticed in two different locations.Indeed, it is observed that the collagen layer has detached from the prosthesis and retracted forming aggregates of collagen.Consequently, there are areas without collagen (non-bright white spots).In this condition, the involved product would have been declared non-compliant in regard to the current list of acceptance and rejection standards applied during the quality control inspection.Considering the product was manipulated, it cannot be excluded that the collagen detachment may be due to product handling.¿.(19/4315) based on the investigation findings, no conclusion can be drawn on the exact origin of the reported collagen peel-off.The result of visual inspection indicates that the cardioroot was handled and cut by the customer.Considering the product was manipulated, it cannot be excluded that the collagen detachment may be due to product handling.Therefore, it is not possible to determine the root cause of the reported event.However, the conducted investigation suggests that the product was not defective at the time of manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOROOT
Common Device Name
PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER
Manufacturer (Section D)
INTERVASCULAR SAS
zone industrielle athelia i
la ciotat 13705
FR  13705
Manufacturer (Section G)
INTERVASCULAR SAS
zone industrielle athelia i
la ciotat 13705
FR   13705
Manufacturer Contact
clemence vaneenoge
zone industrielle athelia i
la ciotat 13705
FR   13705
442084646
MDR Report Key18572740
Report Number1640201-2024-00001
Device Sequence Number7779916
Product Code DSY
UDI-Device Identifier00384401013914
UDI-Public(01)00384401013914
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K103347
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHEWROOT0030
Device Catalogue NumberHEWROOT0030
Device Lot Number23C30
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2024
Supplement Date Received by Manufacturer02/05/2024
Initial Report FDA Received Date01/24/2024
Supplement Report FDA Received Date02/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/30/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-