• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number PXVP2284AT3
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that viscous fluid was found in the reservoir of a truwave vamp during device set up.Hospital staff noticed that a ring in the syringe was left behind from the fluid when pushing the plunger up and back.No patient involvement.
 
Additional Manufacturer Narrative
The device evaluation is anticipated.However, the complaint cannot be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation and device history results when received.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Additional Manufacturer Narrative
A product evaluation of one single dpt vamp plus kit was completed.The reported event of fluid in tubing and vamp was not able to be confirmed.No visible presence of fluid was observed from the kit during pre decontamination examination.Kit was wiped with ipa during decontamination and further examined in product evaluation lab.Microscopic examination of returned kit after decontamination did not detect any presence of fluid inside vamp system and pressure tubing.A device history record review was completed and documented that device met all specifications upon distribution.Further investigation was not required since the reported event was not confirmed and no product defects were detected during product evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE MONITORING KIT
Common Device Name
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer Contact
jonathan diaz
1 edwards way
irvine, CA 92614
MDR Report Key18574525
Report Number2015691-2024-00558
Device Sequence Number20796458
Product Code DXO
UDI-Device Identifier07460691950436
UDI-Public(01)07460691950436(17)251004(11)231005(10)65342602
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K925638
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date10/04/2025
Device Model NumberPXVP2284AT3
Device Lot Number65342602
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2024
Supplement Date Received by Manufacturer04/01/2024
Initial Report FDA Received Date01/24/2024
Supplement Report FDA Received Date04/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-