| Model Number |
PXVP2284AT3 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that viscous fluid was found in the reservoir of a truwave vamp during device set up.Hospital staff noticed that a ring in the syringe was left behind from the fluid when pushing the plunger up and back.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation is anticipated.However, the complaint cannot be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation and device history results when received.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|
| |
|
Additional Manufacturer Narrative
|
|
A product evaluation of one single dpt vamp plus kit was completed.The reported event of fluid in tubing and vamp was not able to be confirmed.No visible presence of fluid was observed from the kit during pre decontamination examination.Kit was wiped with ipa during decontamination and further examined in product evaluation lab.Microscopic examination of returned kit after decontamination did not detect any presence of fluid inside vamp system and pressure tubing.A device history record review was completed and documented that device met all specifications upon distribution.Further investigation was not required since the reported event was not confirmed and no product defects were detected during product evaluation.
|
| |
|