| Medical Device Problem Code |
Deflation Problem (1149)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that it took a considerable amount of time to deflate the balloon.Also, it seems that the balloon of this product inflated asymmetrically.Therefore, the physician discontinued using the device.A replacement device with the same lot number was used to complete the operation.No injury was reported to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The devices was recieved for investigation.The reported problem was confirmed.The lower ligature was observed to have slipped from its intended position which caused the balloon to move past the skive hole and prevented the balloon from deflating.The ligature likely failed due to the pull force on the catheter during the procedure.It is possible that procedural factors including stenosis or calcified plaque caused additional pull force to be used during thrombus removal and contributed to the failure.Our review of the lot history records for this lot did not find any discrepancies either in the manufacturing or packaging process that could be related to this incident.Additionally, there have not been any previous complaints related to this lot number.
|
| |
|