• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFE SPINE DYNA-LINK BARB; INTERVERTEBRAL BODY FUSION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFE SPINE DYNA-LINK BARB; INTERVERTEBRAL BODY FUSION DEVICE Back to Search Results
Model Number 83-005
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The implant was not retrieved by life spine, therefore, without further information the root cause of the reported issue could not be identified.However, it is hypothesized that the reported event may have occured due to the barbs not being appropriately seated inside the pocket of the implant and/or cam lock not being turned using the provided torque handle in the set.Additionally, one contributing factor to the failure could be that the patient had poor bone quality, which might compromise the stability of the barbs in engaging into the bone.
 
Event or Problem Description
It was reported that the barbs had migrated discovered post op.The barbs were removed and screws placed on (b)(6) 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNA-LINK BARB
Common Device Name
INTERVERTEBRAL BODY FUSION DEVICE
Manufacturer (Section D)
LIFE SPINE
13951 quality drive
huntley IL 60142
Manufacturer (Section G)
LIFE SPINE
13951 quality drive
huntley IL 60142
Manufacturer Contact
angela batker
13951 quality drive
huntley, IL 60142
8478846117
MDR Report Key18576502
Report Number3004499989-2024-00001
Device Sequence Number12689848
Product Code OVD
UDI-Device Identifier00190837155131
UDI-Public00190837155131
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K191834
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number83-005
Device Lot NumberEM90195
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/08/2024
Initial Report FDA Received Date01/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age50 YR
Patient SexMale
-
-