• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PERFUSION SYSTEMS MIAR ANTEGRADE ANTEGRADE CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PERFUSION SYSTEMS MIAR ANTEGRADE ANTEGRADE CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO Back to Search Results
Model Number 11014L
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Medtronic received information that during use of this minimally invasive aortic root (miar), it was reported that after going to connect the cardioplegia line to the cannula, it had a small piece of plastic inside of needle.It was reported that the plastic floated upwards with the blood, where it was removed and the needle was rinsed thoroughly with the patient¿s own blood.The customer used the cardioplegia needle for procedure.There was no adverse patient effect reported with this event.Medtronic received additional information that the sterile barrier was not still sealed when the foreign material was identified.Medtronic received additional information that according to the health care professional that was involved in incident, the loose fragment could have embolized into the patient's arterial circulation after the heart-lung machine was started and caused damage and ischemia.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Device evaluation summary:visual inspection of the fm showed evidence of a piece of material.Reason for return was confirmed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIAR ANTEGRADE ANTEGRADE CANNULA
Common Device Name
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18578489
Report Number2184009-2024-00030
Device Sequence Number12689868
Product Code DWF
UDI-Device Identifier00613994495495
UDI-Public00613994495495
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
K100274
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/26/2026
Device Model Number11014L
Device Catalogue Number11014L
Device Lot Number2023041173
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2024
Supplement Date Received by Manufacturer03/14/2024
04/16/2024
Initial Report FDA Received Date01/25/2024
Supplement Report FDA Received Date04/09/2024
04/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-