Model Number 20E
Medical Device Problem Codes Failure to Power Up (1476); Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/03/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Stryker evaluated the customer¿s device and was unable to verify or duplicate the reported issue.Stryker did observe a non-critical issue, in which the device was unable to operate in ac mode, but was able to operate in dc mode with a known good battery.After repairs were made for the non-critical issue, proper device operation was observed through functional and performance testing and the device was returned to the customer for use.
 
Event or Problem Description
The customer contacted stryker to report that their device would not power on.In addition, the device did not recognize paddle when connected.In this state the device would be inoperable and defibrillation therapy would not be available if needed.There was no patient involvement reported with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 20E DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key18579123
Report Number0003015876-2024-00178
Device Sequence Number7922823
Product Code MKJ
UDI-Device Identifier00883873887660
UDI-Public00883873887660
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K130454
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number20E
Device Catalogue Number99507-000106
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/03/2024
Initial Report FDA Received Date01/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/26/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-