| Model Number |
IBAL UKA TIB BEA-RING TRL SZ 4 8MM |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, it was reported by a sales representative via sems (b)(4) that during a case, a tibial bearing trial, ar-601-tbd8, cracked.The trial was replaced and the case was completed with no adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.One unpackaged, ar-601-tbd8, tibial bearing trial, was received for investigation.Visual evaluation noted that the device had cracked.A fragment that had broken off was returned for investigation.Functional testing was not performed due to the damage to the device.The most likely cause is attributed to wear and tear due to repeated use of the device with blunt force.The manufacturing date is 2012.
|
| |
|
Search Alerts/Recalls
|