| Model Number |
STAPLE SEATING PUNCH |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, it was reported by a sales representative via (b)(4) that when the rep twisted the top part to disengage the screw, the tip unevenly clamps down and doesn't close evenly with the driver discovered during a case, no adverse affect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.Complaint is confirmed.Visual inspection identified that the distal tip of the staple seating punch had chipped.Functional testing was not performed due to the tip damage to the device.The most likely cause for the reported failure can be attributed to user error of the device due to interference with the mating instrument.
|
| |
|