| Model Number |
421569 |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Distributor reports "paper on packaging tearing unevenly when opening.He could not quantity how many as he said he opened a few from multiple boxes of all of the lots reported." this emdr covers the unknown number of catheters with packaging defective.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 3005778470.
|
| |
|
Additional Manufacturer Narrative
|
|
According to g905704 ver.23.0 within in-process the peeltest is carried out.Point 5.11.3 peelpack must not be sealed so strong that paper residues remained on the foil or paper tears.Seal should be enough strong to keep paper and foil together without any irregularities.Review of test results recorded was carried out and all samples passed peeltest requirements.Review of the dhr showed that all relevant tests required during the manufacturing and packaging process and final product release had been fulfilled and met the requirements.No nonconformity had been registered during the production and sterilization process of the mentioned lot.(b)(4).The increased trend of complaints related to the pouch paper tears unevenly during opening was observed.Therefore, capa 1832524 has been raised to address the issue.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 3005778470.
|
| |
|
Event or Problem Description
|
|
To date no additional patient or event details have been received.
|
| |
|