| Catalog Number |
CL2000S-RC |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The event occurred on an unspecified date and involved a chemolock¿ w/red cap where the customer reported that the product was attached to a patient when it was noticed that the male and female chemolock adapters came undone and few drops of chemo leaked.The customer reported that when the nurse did an assessment and reattached them, they could easily be pulled back apart, so they were replaced with a new set.There was patient involvement; no harm that was associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A sister sample and photo sample are available for evaluation, however, investigation is not yet complete.
|
| |
|
Additional Manufacturer Narrative
|
|
One photo was shared by the customer, where the item and lot information from the item #cl2000s-rc was observed.No damage or anomalies are confirmed.Sixty-six new samples item #cl2000s-rc, were received for evaluation.As received, based on binomeal stages sampling plan, thirty-five (35) samples were connected using a new chemoport, no disconnections or incomplete connections were observed.Each of the sample was torque test activated confirmed met the product specification and the torque residual as well.All the chemolocks present a good shear tab activation and no damage on the splines.The female luer of the chemolocks were confirmed to comply with iso design expectations complaint of leaks cannot be confirmed or replicated.Lot history review was reviewed and no nonconformities were identified that may have contributed to the reported complaint.
|
| |
|
Search Alerts/Recalls
|