| Lot Number |
SFA00742 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Data Problem (3196)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a cryo-ablation procedure using a smart freeze cryo-console the timer stopped showing on the console.During an ablation the section, below the temperature display, was black instead of displaying the ablation duration.Time and temperature readings also became unavailable on the connected recording system.Ablation was stopped and the console was restarted, and the issue was resolved, it did not occur again during the procedure.The procedure was completed successfully with no patient complications.The console will be retained by the site as the issue was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|