• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO S. DE R.L. DE CV ACHIEVE MAPPING CATHETER - 15 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO S. DE R.L. DE CV ACHIEVE MAPPING CATHETER - 15 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT Back to Search Results
Model Number 990063-015
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Low Blood Pressure/ Hypotension (1914); Cardiac Perforation (2513); Pericardial Effusion (3271)
Date of Event 12/14/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a cryo ablation procedure, after complete right inferior pulmonary vein (ripv) ablation and some right superior pulmonary vein (rspv) ablation, a system notice was received indicating that the safety system detected a compromised outer vacuum.The balloon catheter was replaced which resolved the issue.After three freezes in the rspv and two freezes in the left superior pulmonary vein (lspv), the patient's blood pressure dropped lower than expected.A pericardial effusion was noted on intracardiac echocardiography (ice).The procedure was aborted while the patient was under general anesthesia.The patient's chest was then prepped and accessed; blood was aspirated from the pericardium.The patient was then stabilized and transferred to the operating room where a tear was found on the right carina in the left atrium, between the lspv and left inferior pulmonary vein (lipv).The tear was repaired and the patient was transferred to the intensive care unit (icu) for overnight hospitalization.The patient remains in the icu, is stable, and is "doing well." no further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: product id: afapro28, product type: balloon catheter, product id: afapro28, product type: balloon catheter, product id: 4fc12, product type: sheath, product event summary: the data file was returned and analyzed.One patient file was received and recorded on the reported date of the event.The received patient file showed six applications were performed using the first afapro28 balloon catheter with lot number: 12118.The patient file showed nine applications were performed using a second afapro28 balloon catheter with lot number: 12104.The patient file showed system notice 50032 (the safety system has detected a compromised outer vacuum) during ablation at application number five while the first balloon catheter was in use.In conclusion, the clinical issues (pericardial effusion, atrial tear, and hypotension) occurred during the procedure with no indication that the adverse events were related to the performance or a malfunction of the product and the decision to abort the procedure without use of alternate therapy was based upon the medical judgement of the physician.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE MAPPING CATHETER - 15 MM
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18583099
Report Number9617601-2024-00011
Device Sequence Number8517227
Product Code DRF
UDI-Device Identifier00643169467378
UDI-Public00643169467378
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K102588
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/17/2025
Device Model Number990063-015
Device Catalogue Number990063-015
Device Lot Number227520627
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2024
Initial Report FDA Received Date01/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Life Threatening; Required Intervention; Hospitalization;
Patient Age78 YR
Patient SexMale
Patient Weight102 KG
-
-