| Model Number |
71953-01 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Codes |
Hypoglycemia (1912); Loss of consciousness (2418)
|
| Date of Event |
12/26/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported device has been returned and is currently in process of investigation.A follow-up report will be filed once additional information is obtained.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.The customer indicated this event occurred three times, this report covers 1 of 3 reports.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
A customer reported a display issue with the adc device.The customer observed "frozen" display and was unable to view glucose readings.The customer became hypoglycemic and lost consciousness, requiring treatment of glucagon injection administered by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
|
| |
|
Event or Problem Description
|
|
A customer reported a display issue with the adc device.The customer observed "frozen" display and was unable to view glucose readings.The customer became hypoglycemic and lost consciousness, requiring treatment of glucagon injection administered by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Reader (b)(6) has been returned and investigated.Visual inspection has been performed on the reader and liquid contamination was observed on the usb port.Due to the liquid contamination on usb port the reader was not able to be charged and so the reader did not power on.The reader was further de-cased and liquid contamination was observed due to liquid ingress on the pcba(printed circuit board assembly).Therefore, issue is not confirmed to use.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.The dhrs (device history review) for the libre reader were reviewed and the dhrs showed the libre reader passed all tests prior to release.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Search Alerts/Recalls
|