| Catalog Number |
854M0021 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that during a surgery, three vitality drivers fractured when trying to remove a stuck iliac screw.The screw was ultimately able to be removed and replaced with a shorter screw.This resulted in a 20 minute delay to the procedure but no other patient impact.This is report four of four for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: "device evaluation anticipated, but not yet begun" is erroneous and no longer applies.The device was evaluated.Corrections in g1 and h3.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the tip is twisted and fractured off.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.Reference reports 3012447612-2024-00008 through 3012447612-2024-00011.
|
| |
|
Event or Problem Description
|
|
It was reported that during a surgery, three vitality drivers fractured when trying to remove a stuck iliac screw.The screw was ultimately able to be removed and replaced with a shorter screw.This resulted in a 20 minute delay to the procedure but no other patient impact.This is report four of four for this event.
|
| |
|