| Model Number |
101/870/070 |
| Medical Device Problem Codes |
Inflation Problem (1310); Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Operator of device is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the cuff did not inflate, and even if it did, air would escape when the pilot balloon was pressed.The event occurred during patient use, no patient harm/adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: one sample was received.A visual inspection found the device in good condition.During the functional testing, an inflation test was done.It was found that it is not possible to inflate the sample because it leaks so badly.The leak was identified at the joint of the valve and the pilot balloon.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of product.
|
| |
|