| Catalog Number |
050-87215 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Event or Problem Description
|
|
It was reported by a peritoneal dialysis (pd) patient that a fluid leak took place during pd treatment.During drain 2 of 6 the patient encountered a draining slowly warning.The patient had become disconnected from the stay safe connector and fluid had drained out into the bed.Upon follow up, the patient¿s pd nurse verified there were no symptoms, adverse events, injuries, or required medical intervention as a result of the reported event.The patient indicated that the connector seemed faulty.The patient is using the same cycler and is able to continue pd treatment with no further issues.The cycler set used by the patient is not available for return for physical evaluation by the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported by a peritoneal dialysis (pd) patient that a fluid leak took place during pd treatment.During drain 2 of 6 the patient encountered a draining slowly warning.The patient had become disconnected from the stay safe connector and fluid had drained out into the bed.Upon follow up, the patient¿s pd nurse verified there were no symptoms, adverse events, injuries, or required medical intervention as a result of the reported event.The patient indicated that the connector seemed faulty.The patient is using the same cycler and is able to continue pd treatment with no further issues.The cycler set used by the patient is not available for return for physical evaluation by the manufacturer.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius liberty cycler sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
| |
|