Model Number 100510H2HS-V-A2
Medical Device Problem Codes Break (1069); Stretched (1601); Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was stated to be available for return to the manufacturer for evaluation but has not yet been returned.The alleged product issue could not be confirmed.If the device is received at a later date, an investigation will be performed and a supplemental mdr will be submitted.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with the use of the device.
 
Event or Problem Description
It was reported that during an embolization procedure of a recanalized a-com aneurysm.A coil could not be detached after multiple attempts with two detachment controllers.The doctor decided to remove the coil from the patient and found that the coil had stretched and broke off.Part of the coil was placed in the aneurysm and part in the parent artery.An atlas stent was used to secure the coil wire and the aneurysm was closed.There was no reported patient injury.
 
Additional Manufacturer Narrative
Investigation conclusion: the investigation of the returned coil system found the pusher heater coil stretched, broken and entangled at the middle section, the yellow lead wire broken, and the implant not attached to the pusher with indications of activation observed on the pusher heater coil; however, the pusher's monofilament showed a tensile break shape at the tip, which is consistent with the device experiencing excessive force that exceeded the strength of the monofilament causing the implant to separate from the pusher.As the lead wire was returned broken, this investigation could not test or assess the continuity or resistance of the pusher to determine if a condition existed that would have caused or contributed to the alleged non-detachment issue.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
 
Event or Problem Description
Please see h6 and h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDROSOFT-10-AV
Common Device Name
COILS
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise drive
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise drive
aliso viejo, CA 92656
7142478000
MDR Report Key18592372
Report Number2032493-2024-00084
Device Sequence Number6595398
Product Code HCG
UDI-Device Identifier00816777022998
UDI-Public(01)00816777022998(11)231011(17)280930(10)0000441885
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K080666
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100510H2HS-V-A2
Device Lot Number0000441885
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2024
Supplement Date Received by Manufacturer02/14/2024
Initial Report FDA Received Date01/26/2024
Supplement Report FDA Received Date03/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-