• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC AGBA PICC NAVICURVE: 2L 5.5FR X 55CM BP; CATHETER, INTRAVASCULAR, THERA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC AGBA PICC NAVICURVE: 2L 5.5FR X 55CM BP; CATHETER, INTRAVASCULAR, THERA Back to Search Results
Catalog Number DLX-45552-CURVB
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/02/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that: one side of the peel-away sheath broke off while attempting to separate the two halves.The reported defect was detected during use on patient.The patient condition was reported as "fine".No patient injury/consequence or medical intervention reported.
 
Additional Manufacturer Narrative
Qn#(b)(4) the customer report of a broken peel away sheath tab was confirmed by visual inspection of the returned sample.The customer returned one peel away sheath for analysis.Definite signs of use were observed.Visual analysis revealed that one peel away sheath extrusion was separated directly adjacent to the tab.Remains of the extrusion was still within the separated tab.The other tab was intact.It was also noted that the sheath was entirely peeled down both score lines as intended.The overall length of the sheath measured 2 3/4" via calibrated ruler which was within the specification limits of 2 5/8" - 2 7/8" per the sheath product drawing.The outer and inner diameter of the sheath could not be performed due to the condition of the returned sample.Functional inspection could not be performed due to the condition of the returned sample.A device history record review was performed, and no relevant findings were identified.Based on the customer report, sample received, and comments from manufacturing, the root cause of this event was man ufacturing related.Further investigation has been initiated under teleflex's quality system by the manufacturing site to further investigate this issue.Teleflex will continue to monitor and trend on reports of this nature.
 
Event or Problem Description
It was reported that: one side of the peel-away sheath broke off while attempting to separate the two halves.The reported defect was detected during use on patient.The patient condition was reported as "fine".No patient injury/consequence or medical intervention reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGBA PICC NAVICURVE: 2L 5.5FR X 55CM BP
Common Device Name
CATHETER, INTRAVASCULAR, THERA
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18594371
Report Number9680794-2024-00094
Device Sequence Number6857434
Product Code LJS
UDI-Device Identifier10801902208702
UDI-Public10801902208702
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K220363
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date11/30/2024
Device Catalogue NumberDLX-45552-CURVB
Device Lot Number33F23G0162
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2024
Supplement Date Received by Manufacturer01/30/2024
Initial Report FDA Received Date01/29/2024
Supplement Report FDA Received Date02/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-