| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Prior to a supraventricular tachycardia procedure with the patient prepped, the ensite amplifier did not pass the power-on self-test and the procedure was cancelled.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.The cause for the reported self test issue and subsequent cancellation remains unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
One ensite velocity¿ amplifier was received for evaluation for evaluation.The reported event was able to be confirmed by evaluation of the logs.Based on the information and investigation provided to abbott, the reported event was not able to be duplicated, however the root cause was isolated to an intermittent catheter amplifier board in slot 6.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
|
| |
|