| Model Number |
S404 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/15/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker has depleted from 2.5 years down to 0.5 years battery remaining during the recent checked.Patient is very dependent.Boston scientific technical services (ts) discussed changes in battery could be due to rounding up and down at the checks.As of this time, this device remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker has depleted from 2.5 years down to 0.5 years battery remaining during the recent checked.Patient is very dependent.Boston scientific technical services (ts) discussed changes in battery could be due to rounding up and down at the checks.As of this time, this device remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.Correction to field h6 evaluation conclusion codes.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.Correction to field h6 evaluation conclusion codes.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker has depleted from 2.5 years down to 0.5 years battery remaining during the recent checked.Patient is very dependent.Boston scientific technical services (ts) discussed changes in battery could be due to rounding up and down at the checks.This device was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.Correction to field h6 evaluation conclusion codes.Upon receipt at our post market quality assurance laboratory, detailed mechanical and electrical testing was performed on the device.The device battery status was applicable battery status.The device functioned normally throughout testing.Laboratory analysis determined that this device experienced normal battery depletion; however, it was reported from the field that the estimated longevity remaining value appeared to decrease more quickly than expected between routine follow-ups.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any (even slight) changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please note that, despite the drop in estimated longevity remaining, the actual battery condition did not change significantly between follow-ups.In summary, it was determined that this device experienced normal battery depletion, but declared applicable battery status earlier than previously estimated.Of note, the estimated remaining battery longevity of this pacemaker was designed to be calculated (and trigger corresponding device replacement indicators) based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker has depleted from 2.5 years down to 0.5 years battery remaining during the recent checked.Patient is very dependent.Boston scientific technical services (ts) discussed changes in battery could be due to rounding up and down at the checks.This device was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|