• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL+ UD 36IN 0 S/A CTB-1; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL+ UD 36IN 0 S/A CTB-1; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number VCPB946H
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 1/29/2024.Component code: g07002 device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A photo upon which this medwatch is based has been received, however, the photo evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.Please provide the source or name of person providing answers to follow-up questions (not the person relaying/submitting answers to loc or chu).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Related reports: 2210968-2024-00723, 2210968-2024-00724, 2210968-2024-00725, & 2210968-2024-00726.
 
Event or Problem Description
It was reported a wrong needle pattern on the box.It was reported that the needle pattern on the box should be ethiguard which needle tip should be a little rounded.But actually the needle tip of the pattern is pointed as attached photo shows.This issue was found during goods inspection.Not involved in any surgery.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: (b)(6) 2024 this report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found additional information: d9, h3, h6 h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.The returned sample determined that it was received one sealed box that pertain to product code vcpb946h, lot tdmmbe.Additionally, a photo was provided, and it showed two boxes with different needle tip patterns of the same type of needle cbt-1.Also, it was observed that the boxes belong to different lot numbers (tdmmbe and sgmhmm).Upon visual inspection of the returned box, the information printing on the box of the needle shows an ethiguard needle, sales type ctb-1, and point type blunt tip.In further investigation, it was found that the needle pattern on the box is according to the graphic with version v010 used for the lot tdmmbe.In addition, as per the sampling plan, a visual inspection was performed on thirteen samples.The samples were opened, and the needles belonged to ctb-1.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ UD 36IN 0 S/A CTB-1
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key18596489
Report Number2210968-2024-00727
Device Sequence Number20864101
Product Code GAM
UDI-Device Identifier10705031056725
UDI-Public10705031056725
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K032420
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCPB946H
Device Lot NumberTDMMBE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2024
Supplement Date Received by Manufacturer02/01/2024
Initial Report FDA Received Date01/29/2024
Supplement Report FDA Received Date02/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-