| Model Number |
MAJ-554 |
| Medical Device Problem Code |
Image Display Error/Artifact (1304)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
To date, the device has not been returned to olympus for evaluation.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if additional information is provided by the user facility.
|
| |
|
Event or Problem Description
|
|
It was reported, that the cable head on the video system center was damaged, causing interference when moving it.The issue occurred following the procedure.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the reported event was confirmed.The evaluation found the image with interference is detected showing vertical stripes, when moving the cable.The likely cause of the reported event is due to cable internal failure.Olympus will continue to monitor field performance for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide a correction to b5, d1, d2, d4 model number and udi.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was reported the cable head was damaged, causing interference when moving it.The issue occurred following the procedure.There were no reports of patient harm.
|
| |
|