| Model Number |
A22251C |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported, the roller ball portion of the hf-resection electrode, roller, 24-28 fr., 12° and 30° fell off into the patient during a therapeutic hysteroscopy with dilatation & curettage and endometrial ablation.A vaginal exam was performed to look for and retrieve the roller ball causing prolongation of the procedure for an unspecified duration.The patient's anesthesia was extended by a few minutes.The procedure was completed with a similar device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device referenced in this report has not been returned to olympus.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if additional information is received.This report has been submitted by the importer under this mdr report number 2429304-2024-00093.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to provide an update to field h4.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it was likely that the reported event occurred due to wear and tear.Additionally, the loop at the distal end of the electrode wears out during use and may break, burn or melt.However, the subject device was not returned, and the root cause of the reported event could not be determined.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|