• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30° Back to Search Results
Model Number A22251C
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported, the roller ball portion of the hf-resection electrode, roller, 24-28 fr., 12° and 30° fell off into the patient during a therapeutic hysteroscopy with dilatation & curettage and endometrial ablation.A vaginal exam was performed to look for and retrieve the roller ball causing prolongation of the procedure for an unspecified duration.The patient's anesthesia was extended by a few minutes.The procedure was completed with a similar device.There were no reports of patient harm.
 
Additional Manufacturer Narrative
The suspect device referenced in this report has not been returned to olympus.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if additional information is received.This report has been submitted by the importer under this mdr report number 2429304-2024-00093.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to provide an update to field h4.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it was likely that the reported event occurred due to wear and tear.Additionally, the loop at the distal end of the electrode wears out during use and may break, burn or melt.However, the subject device was not returned, and the root cause of the reported event could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°
Common Device Name
HF-RESECTION ELECTRODE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18596937
Report Number9610773-2024-00274
Device Sequence Number8124681
Product Code GCP
UDI-Device Identifier14042761036815
UDI-Public14042761036815
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K954488
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA22251C
Device Lot Number1000046910
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2024
Supplement Date Received by Manufacturer08/01/2024
Initial Report FDA Received Date01/29/2024
Supplement Report FDA Received Date08/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/04/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-