E1: (b)(6).Firstly, cardinal health contacted liberator from medical supply and they further contacted company.Afterwards receiving this report, the consumer was contacted directly for further information.(b)(6).Based on the available information, this event is deemed to be a reportable malfunction.Batch record review: lot 3g01940 was manufactured on 07/10/2023, in convex 2 pc line, with a total of (b)(4) market units (mkus).Compliance engineer performed a batch record review on 16/jan/2024, to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, system application product (sap) material identification (id) 1161272 and manufacturing order (b)(4).Therefore, no discrepancy related to this issue were found within the documentation.Photograph, video and/or physical sample evaluation: no photographs associated with this case were received and no unused return sample was expected.Conclusion summary of the related event: based on the revision of the observation of the processes involved, interviews to the personnel of the line and the expertise of the triage team, this failure mode was attributed to the following probable causes: machine/equipment machine and process current design.Lifecycle of k-tool mechanical components has not been standardized as part of the machine pm.Method there is not a visual aid or job aid on how to use the qc tooling.There is not a detailed visual aid or job aid on how to perform the mass station set up.There is not a standardize d visual aid for the flange loading stations and certification process for the station.Corrective and preventive actions identified will be taken through corrective action / preventive actions (capa) plan.The investigation associated with related event record has been approved and was complete.No additional action is required, and this complaint will be closed.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092, manufacturing site: 9618003.
|