| Model Number |
MT10-7540 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Nonunion/Delayed-union Bone Fracture (2369)
|
| Date of Event |
01/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of the returned device did not find any indications of wear on the bottom surface or sidewalls of the tulip to indicate that the rod was not properly reduced and contributing to the failure.There have been no other complaints regarding this specific manufacturing batch or lot of pedicle screws and only six total complaints for pedicle screw breakage.As revision surgery was performed to replace the interbody cage after fifteen months of implantation and beach marks or striations were found on the neck of the polyaxial screw, the pedicle screw breakage was likely due to fatigue failure as a result of non-union.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024 pt was taken to surgery for cage removal and psf revision.Original surgery (b)(6) 2022.A tiger shark l was placed today after old cage was removed.During the psf revision, one of the screws was noted to be broken (tulip head was broken from screw shaft) at level s1.
|
| |
|