Model Number VLOCM0335
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the myometrial suturing of a caesarean section, when performing a continuous suturing, the thread broke.Another device was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: d9 (return date), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted one partial suture and one needle.The loop end of the suture was not returned.The needle was returned attached to the suture and parked in foam provided by the account.The suture was returned broken in the barbed section at the base of a barb.The break site was 14 1/8 inches from the needle attachment point.It was reported that the thread broke.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VLOC 90
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18602166
Report Number1219930-2024-00463
Device Sequence Number2845054
Product Code GAM
UDI-Device Identifier10884521803343
UDI-Public10884521803343
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K100257
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2025
Device Model NumberVLOCM0335
Device Catalogue NumberVLOCM0335
Device Lot NumberA2M0203VY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2024
Supplement Date Received by Manufacturer03/25/2024
Initial Report FDA Received Date01/29/2024
Supplement Report FDA Received Date03/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-