|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: e3: reporter is a j&j employee.D9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that this was a posterior cervical fusion (c5 right) for cervical spondylotic myelopathy on jan.11, 2024.In the surgery, symphony was used.After inserting pvfs into c5r, the screw in question and the screwdriver shaft in question are no longer coming off.The screw was removed by moving it slightly and applying a pulling force.The surgeon and the nurse could not remove them outside the body.A spare driver shaft was used and a new screw was opened.The surgery was completed successfully within 30 minutes surgical delay.This report is for one (1) symph nav driver - shaft x20 this is report 2 of 2 for complaint (b)(4).
|