EXACTECH, INC. NV GXL LNR, LIPPED, 36MM ID, GROUP 3 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Model Number |
NV GXL LNR, LIPPED, 36MM ID, GROUP 3 CUPS |
| Medical Device Problem Codes |
Naturally Worn (2988); Lack of Effect (4065)
|
| Health Effect - Clinical Codes |
Loss of Range of Motion (2032); Scar Tissue (2060); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Implant Pain (4561); Insufficient Information (4580)
|
| Date of Event |
01/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: concomitants: (b)(6) - 132-36-53 - nv gxl lnr, lipped, 36mm id, group 3 cups.(b)(6) - 170-36-00 - biolox delta femoral head 36mm od, +0mm.(b)(6) - 186-01-56 - integrip cc, cluster 56mm, g3.(b)(6) - 188-01-08 - wedge plasma x/o sz 8.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
It was reported that a 68 yo male patient, initial right hip implanted on (b)(6), 2017, underwent a revision procedure on (b)(6), 2024, approximately 6 years 2 months post the initial procedure.The surgeon performed a head and liner exchange due to the premature wear of a novation gxl liner.He revised to an xle liner and biolox delta option adapter and head.There were no surgical delays or device breakages during the event.No x-rays were able to be obtained.The patient was last known to be in stable condition following the event.No device returns were available as they were sent to legal.No device images were able to be obtained.No further information.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available information, the patient involved meets the
following risk criteria for early prosthesis wear as specified, implanted with a
component having a shelf age of greater than 2 years.The most likely underlying cause for the revision due to early prosthesis wear reported is a combination of risk factors specified.However, this cannot be confirmed from the reported information and the devices were not available for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: d1/d2a/d2b, h6.Mdr section codes updated/corrected: a, b, c, d, e, g.The most likely underlying cause for the revision reported is prosthesis wear and loss of range of motion and a combination of risk factors: such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed from the reported information and the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|