• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. NV GXL LNR, LIPPED, 36MM ID, GROUP 3 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. NV GXL LNR, LIPPED, 36MM ID, GROUP 3 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number NV GXL LNR, LIPPED, 36MM ID, GROUP 3 CUPS
Medical Device Problem Codes Naturally Worn (2988); Lack of Effect (4065)
Health Effect - Clinical Codes Loss of Range of Motion (2032); Scar Tissue (2060); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Implant Pain (4561); Insufficient Information (4580)
Date of Event 01/05/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: concomitants: (b)(6) - 132-36-53 - nv gxl lnr, lipped, 36mm id, group 3 cups.(b)(6) - 170-36-00 - biolox delta femoral head 36mm od, +0mm.(b)(6) - 186-01-56 - integrip cc, cluster 56mm, g3.(b)(6) - 188-01-08 - wedge plasma x/o sz 8.Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
It was reported that a 68 yo male patient, initial right hip implanted on (b)(6), 2017, underwent a revision procedure on (b)(6), 2024, approximately 6 years 2 months post the initial procedure.The surgeon performed a head and liner exchange due to the premature wear of a novation gxl liner.He revised to an xle liner and biolox delta option adapter and head.There were no surgical delays or device breakages during the event.No x-rays were able to be obtained.The patient was last known to be in stable condition following the event.No device returns were available as they were sent to legal.No device images were able to be obtained.No further information.
 
Additional Manufacturer Narrative
Based on the available information, the patient involved meets the following risk criteria for early prosthesis wear as specified, implanted with a component having a shelf age of greater than 2 years.The most likely underlying cause for the revision due to early prosthesis wear reported is a combination of risk factors specified.However, this cannot be confirmed from the reported information and the devices were not available for evaluation.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: d1/d2a/d2b, h6.Mdr section codes updated/corrected: a, b, c, d, e, g.The most likely underlying cause for the revision reported is prosthesis wear and loss of range of motion and a combination of risk factors: such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed from the reported information and the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LNR, LIPPED, 36MM ID, GROUP 3 CUPS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
miguel sosa
2320 nw 66 court
gainesville, FL 32653
3523164164
MDR Report Key18602639
Report Number1038671-2024-00136
Device Sequence Number2821249
Product Code JDI
UDI-Device Identifier10885862023421
UDI-Public10885862023421
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K070479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/02/2018
Device Model NumberNV GXL LNR, LIPPED, 36MM ID, GROUP 3 CUPS
Device Catalogue Number132-36-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2024
Supplement Date Received by Manufacturer04/15/2024
07/03/2025
Initial Report FDA Received Date01/29/2024
Supplement Report FDA Received Date04/15/2024
07/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/08/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1729-2022
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexMale
Patient Weight96 KG
-
-