• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 700 CX MS PUMP; PROSTHESIS, PENIS, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 700 CX MS PUMP; PROSTHESIS, PENIS, INFLATABLE Back to Search Results
Model Number 72404283
Medical Device Problem Codes Inflation Problem (1310); Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the pump of this inflatable penile prosthesis (ipp) had migrated to the top of the scrotum.This resulted in the device being unable to inflate and the patient being unable to obtain an erection.The physician attempted to reposition the pump in office without success.Surgical intervention is planned.There were no patient complications.
 
Additional Manufacturer Narrative
The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available, the cause that contributed to the reported event cannot be established; a conclusion code of cause not established was assigned to this investigation.
 
Event or Problem Description
It was reported that the pump of this inflatable penile prosthesis (ipp) had migrated to the top of the scrotum.This resulted in the device being unable to inflate and the patient being unable to obtain an erection.The physician attempted to reposition the pump in office without success.Several months later, a revision surgery was performed, during which cylinders and pump were removed and replaced.No additional information was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 CX MS PUMP
Common Device Name
PROSTHESIS, PENIS, INFLATABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18602922
Report Number2124215-2024-04548
Device Sequence Number14164357
Product Code JCW
UDI-Device Identifier00878953003863
UDI-Public(01)00878953003863(17)251109(10)1000484102
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/09/2025
Device Model Number72404283
Device Catalogue Number72404283
Device Lot Number1000484102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/09/2024
Supplement Date Received by Manufacturer07/09/2025
Initial Report FDA Received Date01/29/2024
Supplement Report FDA Received Date07/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age72 YR
Patient SexMale
-
-