• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRAEGER MEDICAL SYSTEMS, INC INFINITY CENTRAL STATION; PHYSIOLOGICAL MONITORING SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRAEGER MEDICAL SYSTEMS, INC INFINITY CENTRAL STATION; PHYSIOLOGICAL MONITORING SYSTEMS Back to Search Results
Catalog Number MS26800
Medical Device Problem Codes No Audible Alarm (1019); Under-Sensing (1661)
Health Effect - Clinical Codes Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the users became aware that the external loudspeakers for the infinity central monitoring station will stay inactive after return of electrical power following an outage or disconnection.There was no patient involvement reported for the particular case.
 
Additional Manufacturer Narrative
A followup report will be submitted upon completion of this investigation.
 
Additional Manufacturer Narrative
In this case, the issue was noted during installation of a black box system/kvma (keyboard, video, mouse, audio) extender with a new edifier speaker (using an analog connection).The infinity central station (ics) instructions for use (ifu) warns users that external speakers must always remain connected and powered on and to perform regular function checks when using kvma extenders.Investigation confirmed that when the new edifier speakers [r19umkii] are used in analog mode, the speakers will not function if the analog cable is connected (into the speaker rear aux jack) before the speaker is powered up (initially or after loss of power).After a temporary loss of power, the analog connection must be reset by unplugging and then re-plugging the analog cable into the speaker rear aux jack.The issue occurs because the default speaker connection mode is usb.The usb connection is required for a power source even when used in analog mode.The mode is set to analog by connecting the analog cable into the speaker rear aux jack after the speaker is powered up.The speaker mode is indicated by the led next to the power button: the led is green for analog connection mode; the led is red for usb connection mode.
 
Event or Problem Description
It was reported that the users became aware that the external loudspeakers for the infinity central monitoring station will stay inactive after return of electrical power following an outage or disconnection.There was no patient involvement reported for the particular case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY CENTRAL STATION
Common Device Name
PHYSIOLOGICAL MONITORING SYSTEMS
Manufacturer (Section D)
DRAEGER MEDICAL SYSTEMS, INC
3135 quarry road
telford PA
Manufacturer (Section G)
DRAEGER MEDICAL SYSTEMS, INC
3135 quarry road
telford PA 18969
Manufacturer Contact
3135 quarry road
telford, PA 18969
2156602400
MDR Report Key18604341
Report Number1220063-2024-00012
Device Sequence Number2821295
Product Code MHX
UDI-Device Identifier04049098001878
UDI-Public(01)04049098001878(11)140226(93)MS26800-09
Combination Product (Y/N)N
PMA/510(K) Number
K061379
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Biomedical Engineer
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 12/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMS26800
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/19/2024
Supplement Date Received by Manufacturer12/30/2024
Initial Report FDA Received Date01/30/2024
Supplement Report FDA Received Date12/30/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/26/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
1220063-12232024-001-C
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NA.; NA.
Patient SexUnknown
-
-