• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH BIOPSY FORCEPS, SPOON TYPE, OPTICAL, FOR 12° TELESCOPE; RIGID ENDOTHEREAPY BIOPSY FORCEPS, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH BIOPSY FORCEPS, SPOON TYPE, OPTICAL, FOR 12° TELESCOPE; RIGID ENDOTHEREAPY BIOPSY FORCEPS, REUSABLE Back to Search Results
Model Number A20712A
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/14/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for evaluation, and the evaluation found that the screw at the base of the handle had come loose.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or when additional information becomes available.
 
Event or Problem Description
It was reported that the biopsy forceps's tip did not open and close when the handle was opened and closed.The autoclave was used as specified.The issue was found during preparation for use.There were no reports of patient harm.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 1 years since the subject device was manufactured.The device was returned to olympus for inspection, and the customer's allegation that the tip did not open and close when the handle was opened and closed was confirmed.Based on the results of the investigation, it was likely that the event was caused by wear and tear, poor maintenance, or improper handling.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOPSY FORCEPS, SPOON TYPE, OPTICAL, FOR 12° TELESCOPE
Common Device Name
RIGID ENDOTHEREAPY BIOPSY FORCEPS, REUSABLE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg/de/germany 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg/de/germany
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18605398
Report Number9610773-2024-00285
Device Sequence Number15735293
Product Code GCP
UDI-Device Identifier04042761020008
UDI-Public04042761020008
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K790071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA20712A
Device Lot Number22702-0011
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/14/2024
Supplement Date Received by Manufacturer04/30/2024
Initial Report FDA Received Date01/30/2024
Supplement Report FDA Received Date05/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/15/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-