| Model Number |
865040 |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported a speaker malfunction.It is unknown if the device was in clinical use at the time of the event.No adverse event or patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A philips remote service engineer (rse) spoke by telephone support with the customer biomed and assisted with functional testing of the device.Functional testing results confirmed the speaker malfunction inop with no sound being emitted.The rse determined the speaker would require replacement.The customer biomed replaced the speaker to resolve the reported issue and the device remains at the customer site.(b)(6).
|
| |
|