| Model Number |
10094137 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Siemens became aware of a malfunction that occurred while operating the artis zee ceiling system.During an emergency patient procedure, the unit turned off, which resulted in a treatment delay.We have no indications of any adverse effects on the health status of the involved patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens is conducting a thorough investigation of the reported event.A supplemental report will be filed upon completion of the investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation of the reported event was completed by our experts.It showed an issue with the digital image preprocessing (dipp).The dipp reset by itself but the connection to the flat detector (fd) could not be re-established.According to the expert opinion, a possible reason for this issue was either a defective flat panel detector receiver (fdr) board, a component within the dipp, or a bad connection between the fdr and the fd.Following the replacement of the fdr board, the issue was resolved.The occurrence rate of the aforementioned error pattern was checked.A possible error accumulation or even a systematic error, which leads to a corrective action of the installed base, could not be determined by the investigation.
|
| |
|