• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS PATIENT INFORMATION CENTER IX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS PATIENT INFORMATION CENTER IX Back to Search Results
Model Number 866389
Medical Device Problem Code Alarm Not Visible (1022)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2024
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the device presented that the battery alarm on cardio telemetry missing.It is unknown if the device was in use at time of event, there was no adverse event reported.
 
Additional Manufacturer Narrative
Philips is in the process of obtaining additional information regarding the reported event and the investigation is ongoing.A follow-up report will be submitted upon completion of the investigation.(b)(6).
 
Additional Manufacturer Narrative
The remote service engineer (rse) reviewed the log files provided by the customer, but the log files received did not contain the time in question.The rse contacted the customer again to request the audit log files that contained the time in question.The client stated that they did not wish to investigate further and did not provide the logs.Philips cannot rule out a malfunction of the device, as provided log files from the customer did not contain for the reported time of event, which was needed to perform a full investigation, was not provided by the customer and the exact cause for the reported issue remains unknown.
 
Event or Problem Description
The device was in use at time of event, there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATIENT INFORMATION CENTER IX
Common Device Name
PATIENT INFORMATION CENTER IX
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18606797
Report Number1218950-2024-00057
Device Sequence Number14158423
Product Code MHX
UDI-Device Identifier00884838104594
UDI-Public884838104594
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K211900
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number866389
Device Catalogue Number866389
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2024
Supplement Date Received by Manufacturer01/29/2024
Initial Report FDA Received Date01/30/2024
Supplement Report FDA Received Date02/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-