• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; Stimulator, electrical, implanted, for parkinsonian tremor

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; Stimulator, electrical, implanted, for parkinsonian tremor Back to Search Results
Model Number 37601
Medical Device Problem Codes Corroded (1131); Failure to Power Up (1476); Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the battery compartment was corroded and the patient programmer won't turn on.The patient couldn't see the serial number.A replacement programmer was sent out.No symptoms were reported.
 
Additional Manufacturer Narrative
Concomitant medical product: product id 37642 serial# unknown product type programmer, patient section d information references the main component of the system.Other relevant device(s) are: product id: 37642, serial/lot #: unknown, h3: analysis of the 37642 programmer (s/n unknown) revealed that corrosion was found on the boards and the device was scrapped.This regulatory report is being submitted as part of a retrospective review as part of a remediation plan 411 action.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Common Device Name
Stimulator, electrical, implanted, for parkinsonian tremor
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18607546
Report Number3004209178-2024-02720
Device Sequence Number11474272
Product Code MHY
UDI-Device Identifier00643169100770
UDI-Public00643169100770
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P960009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/14/2019
Device Model Number37601
Device Catalogue Number37601
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2021
Initial Report FDA Received Date01/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
-
-