Model Number DTBB1D4
Medical Device Problem Code Pacing Problem (1439)
Health Effect - Clinical Codes Bradycardia (1751); Endocarditis (1834); Sepsis (2067); Heart Failure/Congestive Heart Failure (4446)
Date of Event 12/15/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continuation of d10: 620bg27 heart band medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient was admitted to the hospital due to heart failure with a septic blood infection.The source of the infection was bacterial endocarditis.The sepsis was treated with six weeks of antibiotics.At the time of admission, a dialysis port was placed through the same vein the leads were placed in ¿ the right subclavian vein.The physician believed this damaged the right atrial (ra) lead, causing oversensing and undersensing.Small p waves or noise caused the cardiac resynchronization therapy defibrillator (crt-d) to not properly pace and a loss of crt pacing occurred resulting in bradycardia.This further lead to heart failure, fluid overload, and an irregular heart rhythm resulting in short-long-short episodes that were proarrhythmic.The patient stopped their antiarrhythmic and high blood pressure medications, and their sodium and potassium levels were out of range, thus, making pro-arrythmia more likely.Approximately four weeks after the hospital admission, the ra lead was reprogrammed leading to one hundred percent crt pacing.The device and lead remain in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was further reported that both the atrial and right ventricular (rv) leads had vegetation.The device, the atrial, rv and left ventricular (lv) leads were explanted and replaced one week later due to the endocarditis infection.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA¿ S CRT-D
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18607718
Report Number3004209178-2024-02727
Device Sequence Number9078076
Product Code NIK
UDI-Device Identifier00643169007178
UDI-Public00643169007178
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P010031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/14/2017
Device Model NumberDTBB1D4
Device Catalogue NumberDTBB1D4
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2024
Supplement Date Received by Manufacturer03/08/2024
04/26/2024
Initial Report FDA Received Date01/30/2024
Supplement Report FDA Received Date03/26/2024
05/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/23/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-45 LEAD, 419688 LEAD, 6947M55 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Life Threatening;
Patient Age73 YR
Patient SexMale
Patient Weight108 KG
-
-