| Catalog Number |
09005803190 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The analyzer serial number is (b)(6).The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys ft3 iii results for 2 patient samples on a cobas 6000 e 601 module.The customer questioned the high initial and repeat results as they did not match the patients' clinical histories and was prompted to change the reagent pack and repeat the samples again.For patient 1, the initial ft3 result was 12.8 pmol/l.The sample was auto-repeated and the result was 12.9 pmol/l.The sample was repeated again with a different reagent pack and the result was estimated to be 2-3 pmol/l.For patient 2, the initial ft3 result was 10.8 pmol/l.The sample was auto-repeated and the result was 11.95 pmol/l.The sample was repeated again with a different reagent pack and the result was estimated to be 2-3 pmol/l.No questionable results from patient 2 were reported outside of the laboratory.The estimated results of 2-3 pmol/l were deemed correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The qc recovery data provided was acceptable.The alarm trace showed sample short and abnormal probe sucking alarms.The customer replacing the reagent pack resolved the issue.No further issues were reported.
|
| |
|