| Catalog Number |
46100-91 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The event involved a cath lab w/5 port "off" manifold (600 psi), 72" admin set and 4 ft transpac® iv.It was reported there was a set leaking and connections were not secure.The products were noted to have transducers which would not stay secured to 5 port manifold and fluid would leak around having to switch out the transducers for new there was patient involvement and no adverse event or human harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation; however, testing has not yet been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported complaint of leakage and connections were not secured was not confirmed.During visual inspection, one of the manifold handle was received broken.It is unknown how and when the manifold handle was broken.The probable cause of the broken handle is unknown.The returned set was primed and pressure leak tested and no leaks were observed.All the connections secured well.The lot history was reviewed and no nonconformities were identified that may have contributed to the reported complaint.
|
| |
|