• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. CATH LAB W/5 PORT "OFF" MANIFOLD (600 PSI), 72" ADMIN SET AND 4 FT TRANSPAC® IV; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. CATH LAB W/5 PORT "OFF" MANIFOLD (600 PSI), 72" ADMIN SET AND 4 FT TRANSPAC® IV; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number 46100-91
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/11/2023
Type of Reportable Event Malfunction
Event or Problem Description
The event involved a cath lab w/5 port "off" manifold (600 psi), 72" admin set and 4 ft transpac® iv.It was reported there was a set leaking and connections were not secure.The products were noted to have transducers which would not stay secured to 5 port manifold and fluid would leak around having to switch out the transducers for new there was patient involvement and no adverse event or human harm.
 
Additional Manufacturer Narrative
The device was returned for evaluation; however, testing has not yet been completed.
 
Additional Manufacturer Narrative
The reported complaint of leakage and connections were not secured was not confirmed.During visual inspection, one of the manifold handle was received broken.It is unknown how and when the manifold handle was broken.The probable cause of the broken handle is unknown.The returned set was primed and pressure leak tested and no leaks were observed.All the connections secured well.The lot history was reviewed and no nonconformities were identified that may have contributed to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATH LAB W/5 PORT "OFF" MANIFOLD (600 PSI), 72" ADMIN SET AND 4 FT TRANSPAC® IV
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18608619
Report Number9617594-2024-00064
Device Sequence Number15735317
Product Code DQO
UDI-Device Identifier00840619070601
UDI-Public(01)00840619070601(17)260201(10)13538406
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K932141
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2026
Device Catalogue Number46100-91
Device Lot Number13538406
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2024
Supplement Date Received by Manufacturer02/01/2024
Initial Report FDA Received Date01/30/2024
Supplement Report FDA Received Date02/29/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-