| Model Number |
0998-UC-3023-53 |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) the cs300 intra-aortic balloon pump (iabp) the display had horizontal lines on the bottom as well as blurry spots.There was no patient involvement reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The service territory manager (stm) evaluated the unit on a later date.The stm successfully completed a function check with testing catheter and trainer without issue upon initially testing unit.In order to solve the issue, the stm replaced display, display grid/casing, and display cable, and successfully obtained a clear display without issues.The coiled cord, when attempted out of the box, produced electrical test fails code 120.The existing coiled cord was left on unit, fully functional, which would have been replaced as a precaution.Preventative maintenance (pm) was completed successfully with measurements and calibration information separately.The unit was calibrated and passed all functional and safety tests per factory specifications.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: b4, g3, g6, h2, h10, h11 corrected fields: d1, d4.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: b4, g3, g6, h2, h6, h11.Corrected fields: d4, h6 (type of investigation, investigation findings, investigation conclusion).The failure analysis and testing dept.Received the following parts associated with this complaint: part number (b)(4) coiled cord serial number (b)(6) part number (b)(4) mounting plate serial number na part number (b)(4) cable to video receiver serial number 20 26_meritec part number (b)(4) display serial number (b)(6) these parts were received with a reported unit failure of the display having horizontal lines on the bottom as well as blurry spots.The failure analysis and testing dept.Performed a visual inspection and found the parts to be in good condition, except part number (b)(4) display serial number (b)(6), that had scratches and marks on it.The failure analysis and testing dept.Installed the parts into the cs300 test fixture and tested the parts to factory specifications per the cs300 service manual.The fat could not verify the failure of horizontal lines on the bottom of the display, and blurry spots after an extended run-time of the cs300.All parts passed testing.Retaining the parts in the fat per procedure.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates only.Providing updated device identification information in alignment with gudid (d1, d2, d3, d4, g4, h4, h5).
|
| |
|